Labotect/CO? Incubator C16/CO? Incubator C16/1 Ea
温控控制
商品编号
CO? Incubator C16
品牌
Labotect
公司
LABOTECT GMBH
公司分类
CO? Incubators
Size
1 Ea
商品信息
REF 14209
(without O? control)
REF 14452
(with O? control)
Incubator with an inner volume of 16 liters
Advantages of C16:
Approved medical device
CO? sensor:
Dual beam infrared sensor
Very precise measurement
Very short recovery times
Long-term stable
Direct Heating:
Independent over-all heating of inner walls and door
Homogeneous temperature distribution in the interior
Very short recovery times?
Very compact and space saving construction
?
C16_english_5_012017.pdf
Features
Technical data
Options
Interior
Electro-polished stainless steel with coved corners
Shelves and ductwork easily removable for cleaning purposes
Temperature
Panel heated interior chamber and door
Humidity supply
Internal water reservoir
CO?
Dual beam infrared sensor
Measurements independent from temperature and humidity
Very short recovery times
For all adjustable parameters through optimized microprocessor control
Easy Handling
Menu guided
Backlit LCD display
Foil keypad
Diagnostic system
Optical and acoustical alarm if set point deviation or sensor defect occurs
Door monitoring
Air monitoring
Hardware self-test
Remote alarm
Options
O? control: connection for N?
Interfaces
Remote alarm (potential free)
RS485-interface for data-logging with optional PC software LaboDat+
Exterior
Dimensions 290 x 330 x 480?mm (w?x?d?x?h)
Interior
Volume 16?l
Electro-polished stainless steel
2 shelves 205 x 230?mm (w?x?d)
Temperature
Panel heated interior chamber and door
Range 27?°C - 42?°C starting 5?°C above ambient temperature
St
ABI
lity / uniformity: ±?0.1?°C?/?±?0.3?°C
CO?-control
Dual beam infrared sensor
Range 0 – 10 % CO?
St
ABI
lity ± 0.1 % CO?
O?-control < 21 % (optional)
Measurement with galvanic sensor
Connection of N?
Range 1 % – 21 % O?
Supply
110 – 230 V AC, 50/60 Hz, 60 W
Inlet pressure for gas 0.8 bar
Ambient temperature 18?°C?-?30?°C
Classification
Safety class I
Class IIa for every usage according to EC-Directive 93/42/EEC
EN 61010 conform
Our incubator C16 is available with O
2
control as option (O
2
< 21 %)
Further accessories for incubator C16
Gas In-Line-Filter
PC software LaboDat+
产品货号:待定